AJP - GI Fuel your research with LabChart
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Am J Physiol Gastrointest Liver Physiol 279: G837-G843, 2000;
0193-1857/00 $5.00
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Craig, A. G.
Right arrow Articles by Dent, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Craig, A. G.
Right arrow Articles by Dent, J.
Vol. 279, Issue 4, G837-G843, October 2000

SPECIAL COMMUNICATION
Evaluation of multiple-point measurement of sphincter of Oddi motility in the Australian brush-tailed possum

A. G. Craig1, T. I. Omari2, G. T. P. Saccone1, J. Toouli1, and J. Dent3

1 Department of General and Digestive Surgery, Flinders Medical Centre, Bedford Park, South Australia 5042; 2 Gastroenterology Unit, Women's and Children's Hospital, North Adelaide, South Australia 5000; and 3 Department of Gastrointestinal Medicine, Royal Adelaide Hospital, Adelaide, South Australia 5000, Australia

Manometric assembly diameter is a major limitation on the number of perfused manometric recording points for recordings from the sphincter of Oddi (SO). We evaluated novel polyimide manometric assemblies whereby four recording channels were incorporated in an overall assembly diameter of 0.8 mm. Over the very low range of perfusion rates tested (0.005-0.04 ml/min), the assemblies had pressure offsets attributable to water perfusion from 2 to 23 mmHg and pressure rise rates from 20 to 163 mmHg/s. In six anesthetized Australian brush-tailed possums, manometric recordings from the SO showed a significant reduction in the recorded peak amplitude of pressure waves with perfusion rates below 0.02 ml/min. The pressure profile of the sphincter was found to be asymmetric, and phasic wave propagation patterns were complex (antegrade 35.6%, "mixed" 64.4%). In conclusion, accurate multipoint SO manometry in the possum can be performed with micromanometric assemblies at very low perfusion rates to give a more complete understanding of SO mechanics. These methods are also potentially applicable to perfusion manometry in other small laboratory animals such as mice.

perfusion manometry; micromanometry; recording fidelity


This article has been cited by other articles:


Home page
J. Physiol.Home page
M. B. Grivell, C. M. Woods, A. R. Grivell, T. O. Neild, A. G. Craig, J. Toouli, and G. T. P. Saccone
The possum sphincter of Oddi pumps or resists flow depending on common bile duct pressure: a multilumen manometry study
J. Physiol., July 15, 2004; 558(2): 611 - 622.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Visit Other APS Journals Online